SMV-2015-23: Development of a low temperature system for a spectroscopy
Urban, Pavel ; Hanzelka, Pavel
The system for measurement of temperature of solid and liquid samples for spectroscopic measurement in the range of 77 K - 300 K was designed. For the above mentioned system, the pump vacuum line was designed and analysed for pumping of vacuum spaces of two types of liquid nitrogen cryostats used for spectroscopic measurements. The part of the solution is design of temperature sensor positions and an electrical resistance heater for the sample temperature stabilisation. Furthermore, an electrical resistance heater for heating/outgassing of the vacuum pump line was designed.